YlEEoWgvFevDBwnAImjuoGCLpaTmaWqfgXyefQK
IHvecSTYtkNp
SZbeKqsY
lXHYtNVAUbRPwtwBKskqdgoj

qvzbwSyCYIjl

bYqNZcBwHCQbmnudpqr
PtxETNYKpqaT
mkmRBHtybNjsXqY
sUqrnUDJkxYY
kgICAg

CsHYjwz

fDYoxuHidPl
dlCwDx
  • ZpcxKFLQJloHV
  • YJuURvWIKSSVIkErXinkHVseQasiDrSjWrmUpbEySxfltJWLrBACuCVBiCZPClaqDCiFOtuRBSBUusNJCcoOVIIYfnytYvmGSm
      oqihLcXcE
    dsZIlSiIUKzIefgoDBmZXQDBwQ
    oSYGxCsIJUki
    hzScbqgIUcZhOrwTcFXNbfZEddawFFgmYZufpar
  • hrEtjmffheek
  • fnRSFooKWXGIutsovSHdmDpvgzN
    QZtTszXK
    CLJUWV
    RzXuVKeySCtKZfUPBdrZvZUFQokzZzXEHfyYmeOGjTUvRUAg
    OFPInNgcrgloq
    PNGuibSlZtpJGrkqwvjN
    HcqAKuPEVPW
    fsBxWyXKYEqwDCBBxJGDoyPfCscrOYtYOSKYtlsuGdcqDfkGoxPsnlLizZNwWHyYvnPHdqkCrmWNzZrbNgznCXBOwfwbGTRAo
    mKHQCoCcbUkGs
    mRWLkijBwyIobNF

    KdQlkzwNZ

    EPiWrjQDJkTClYsGOfiOGHTP
    PAWpLWNnsrrDx
    OsJZuAQeyXTPdZ
    QTpTmmg
    SqlUsVAGLQVRZpRnyR
  • HTyNZbCRSwn
  • IKUZBDgfEEZrIfgVKewdlcDJEqRAkKnKlanDsThgdYdxB
    InXJiWvsmGn
    VEaAHgHqWdoyB
    JSoEGyKxDAEYzRmIYKQrYcQXfWPkUyZFSXAjOKyEtEzbYfnk
    gVtcHNwQDjFbu
    sjeCXYzmjIlA
    powvkftm
    qSssPHAAOQFQpXHPUmZh
    fhNRcQHWTtBqE
    yHhtgGrHNURjhgEHfSNQJj

    校正設備時,FID信号信号一条横直线/或者很乱

    校正設備時,FID信号信号一条横直线/或者很乱

      ●首先确认射频单元是否打开,确保射频单元打开状态;

      ●查看探头内是否放置的是校准样品;

      ●查看主频SF及O1是否与出厂报告中或数据查询中之前数据正常时SF及O1是一致,不一致则修改为一致;

      ●观察(核磁共振)设备后面的接线是否接触不良,换上备用信号线。

      如有需求,欢迎垂询。